pH, Turbidity PHotoFlex series
Product Descriptions
pHotoFlex® Turb : portable multi-parameter photometer with full value pH, ORP and turbidity measurement for water analytics and environmental monitoring
– Versatile with special methods NH3 and CO2
– Unique combination with pH and turbidity measurement
– Turbidity measurement of 0.01 – 1,100 NTU according to DIN ISO 27027
– Light source : LED
– Wavelength nm : 436, 517, 557, 594, 610, 690 + 860
– Own methods : 100
– Data memory : 1,000 measurements
– pH/Redox : pH 0-16 with automatic temperature control (ATC)
– Turbidity : 0-1100 NTU/FNU, nephelometric, DIN ISO conform
– Accuracy : Photometry: <2 nm Wellenlängengenauigkeit, 0,005 Abs Reproduzierbarkeit
– pH: ±0.01 pH
– Turbidity (NTU / FNU):0.01 NTU/FNU or ±2% of the measured value
The energy-efficient LED optics with 6 wavelengths and a smart adapter for a variety of applications with 16 and 28 mm cuvettes allows the measurement of the smallest concentration with more than 150 programs for standard parameters, special methods and coloration. There is a large selection of test sets of lot-certified cuvettes all the way to economic powder reagents available. A list of favorites for the most commonly used tests, sample ID numbers and more than 3,000 measurements per battery set facilitate field measurements.
The integrated pH function is a complete electro-chemical pH measurement of pH 0-16 with automatic buffer recognition (TEC/NIST) by using the comprehensive WTW selection of pH measuring chains. The temperature compensation automatically takes place in the admissible measuring ranges from -5 … 100 °C. The WTW MultiCal® routine allows the automatic calibration with up to 3 calibration points. or field use, there is, for instance, the maintenance-free SenTix® 41 is recommended, for precision measurements in the lab, the glass electrode SenTix® 81 is frequently used.
The pHotoFlex®Turb also has an infrared light source (IR) for the nephelometric turbidity measurement (90°) according to DIN 27027 / ISO 7027. In combination with th e long-term stable AMCO Clear® standard, the highest possible precision is also warranted for the sensitive drinking water area.
pHoto Flex | STD | pH | Turb |
Order-Nr. pHotoFlex® Instrument (SET) | 251 105 (251 205) | 251 100 (251 200) | 251 110 (251 210) |
Photometric Measurement | Transmission (%T), Absorption, Concentration | ||
pH-Measurement (±0.01 pH)/ Redox | – | / | / |
Turbidity Measurement IR acc. DIN /ISO | – | – | / |
Calibration Intervall /-protocol | – | / | / |
Programs | more than 150 | ||
Special programs with pH and Temperature | – | NH3, CO2 | |
User-defined programs via LSdata Software | 50 | 100 | |
Data Sets (with Identification number ID) | 100 | 1000 | |
Field Sets with Accessories | optional | ||
Optics | power saving LED + optical Filters: 436, 517, 557, 594, 610, 690 nm 0.005 Abs reproduceability, < 2 nm wavelength accuracy |
||
Cell sizes | 16 mm round ( H:91-104 mm); 28 mm | ||
Batteries | 4 AA for approx. 3000 measurements | ||
Interface, Update via Internet | RS232 + cable AK 540 B (Accessories) | ||
LSdata PC Software | included in Sets or indivudal package optional | ||
Lab use: optional Accessories | LabStation with barcode option via Scanner, incl. recharg. batteries and PC software LSdata, cable | ||
Recharg. batteries + universal power supply | – | optional | optional |
Certificates | CE/ETLus/cETL |